Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can a Leica DISTO ™ D3 be calibrated by the user?

calibrated D3 disto leica user
0
Posted

Can a Leica DISTO ™ D3 be calibrated by the user?

0

The user can calibrate the device’s tilt sensor. The angle sensor is switched on by pressing the function key once. Then two measurements are taken on a flat surface. The first measurement is taken and a note is made of the measured angle a. The device is then turned through exactly 180o, the second measurement is taken and a note is made of the measured angle b. The value x to which the device must be corrected is calculated as follows: x = -(a+b) /2 The calibration mode is then entered by pressing keys “Clear Off” and the “Function key” both at the same time for 2 seconds. The correction value x can be entered here using the “plus” and “minus” keys. The entered value is stored and implemented by pressing the “Menu/=” key.