Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can <30 or >450 be entered as appropriate values for the PCx glucose test?

appropriate Glucose Test values
0
Posted

Can <30 or >450 be entered as appropriate values for the PCx glucose test?

0

• Yes. However, exceptions cannot be recorded when the result value is non-numeric. If a patient ID exception (or other exception) were to occur while processing one of these values, the exception would not be logged. Note that the presence of “<" or ">” signs in a result value does not cause an exception. To enable exception logging when the result value is non-numeric, it is necessary to modify the RESULT field definition in the PRECISION G EXCEPTION file. Use Fileman “Modify File Attributes” option: MODIFY WHAT FILE: PRECISION G EXCEPTION// Select FIELD: 1 INSTRUMENT SERIAL NUMBER (multiple) LABEL: INSTRUMENT SERIAL NUMBER// ^ Select INSTRUMENT SERIAL NUMBER SUB-FIELD: 1 PATIENT I.D. (multiple) LABEL: PATIENT I.D.// ^ Select PATIENT I.D. SUB-FIELD: RESULT VALUE LABEL: RESULT VALUE// ^ DATA TYPE OF RESULT VALUE: NUMERIC// FREE TEXT MINIMUM LENGTH: 1 MAXIMUM LENGTH: 4 (OPTIONAL) PATTERN MATCH (IN ‘X’): IS RESULT VALUE ENTRY MANDATORY (Y/N): NO// NO …. ‘HELP’-PROMPT: Answer must be 1