are we sure that high-performance liquid chromatography currently works in the clinical routine?
The clinical usefulness of glycated hemoglobin (HbA1C) depends crucially on the accuracy and precision of its assay. When we compared an immunological bench-top analyzer (DCA 2000, Bayer Diagnostici, Milan) to the high-performance liquid chromatography (HPLC) reference method used in a routine hospital laboratory (Diamat and Fast Diamat, Bio-Rad Lab., Milan) by assaying multiple control sera, we found so many sources of systematic analytical errors in the routine use of HPLC as to compromise between-assay precision. DCA 2000 showed intra- and interassay coefficients of variation (CV) of 1.1% and 2.3% with the normal standard serum, 1.0% and 4.2% with the pathological one; Diamat yielded CVs of 1.3% and 7.0%, 1.3% and 5.7%, respectively. Although the measurement of 161 blood samples showed that Diamat usually overestimated HbA1C (paired t-test, P<0.001), a great variability of Diamat performance became evident when the relationship Diamat vs DCA was evaluated day by day over 17 days of