Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Are there interferences when testing with CoaguChek systems?

0
Posted

Are there interferences when testing with CoaguChek systems?

0

• Polyglobulism: Hematocrit outside the standard range (25 – 55 %) • Lupus anticoagulant: INR abnormally high The presence of anti-phospholipid antibodies (APAs) such as Lupus antibodies (LA) can potentially lead to prolonged clotting times, i.e. elevated INR values. A comparison to an APA-insensitive laboratory method is recommended if the presence of APAs is known or suspected.