Are self-expanding stents superior to balloon-expanded in dilating aortas?
OBJECTIVE: To study the stent/vessel interaction and distensibility following the natural increase in vascular diameter using self-expanding and balloon-expanded stents. DESIGN: Open experimental study. SETTING: Animal laboratory, university hospital. MATERIALS AND METHODS: Eight Palmaz (P) and eight Gianturco (G) stents were transluminally placed in the infrarenal aortas of 16 pigs. Pulsatile diameter changes above, at and below the stents were non-invasively monitored with an ultrasound phase-locked echo-tracking system before and immediately after stenting and at 4 and 18 weeks. Blood pressure was registered intra-arterially and stiffness (beta) was calculated. Intravascular ultrasound (IVUS) was performed at 18 weeks. RESULTS: Median weight increased from 20 kg (19-26) to 93 kg (62-130). Diameter of the aorta increased 60%. In group P no pulsatile diameter change could be measured at the stent (beta = infinity). In group G stenting increased stiffness from beta 20.7 (9.2) to 43.2 (