Are routine arrhythmia inductions necessary in patients with pectoral implantable cardioverter defibrillators?
INTRODUCTION: The value of ventricular arrhythmia inductions as part of routine implantable cardioverter defibrillator (ICD) follow-up in new-generation pectoral ICDs is unknown. METHODS AND RESULTS: We performed a retrospective analysis of a prospectively collected database analyzing data from 153 patients with pectoral ICDs who had routine arrhythmia inductions at predismissal, and 3 months and 1 year after implantation. Routine predismissal ventricular fibrillation (VF) induction yielded important findings in 8.8% of patients, all in patients with implantation defibrillation threshold (DFT) > or = 15 J or with concomitant pacemaker systems. At 3 months and 1 year, routine VF induction yielded important findings in 5.9% and 3.8% of tested patients, respectively, all in patients who had high DFT on prior testing. Ventricular tachycardia (VT) induction at predismissal, and 3 months and 1 year after implantation resulted in programming change in 37.4%, 28.1%, and 13.8% of tested patient