Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Allele Calling Software 27. What can you suggest for setting up the instrument(s) and or measurement of FP?

0
Posted

Allele Calling Software 27. What can you suggest for setting up the instrument(s) and or measurement of FP?

0

• Proper filters must be installed in the instrument to measure the fluorescence of R110 (Ex 490 nm/Em 520 nm max) and TAMRA (Ex 550 nm/Em 570 nm max). Some instruments also require that the appropriate dichroic mirrors be installed. • For the Victor2 V the recommended filters are: • R110 485ex 535/25em • TAMRA 544ex 595/60em • Setting or adjusting of the “G” factor is highly recommended to facilitate troubleshooting and comparisons at different sites with different instruments. The G factor is a calibration that sets the mP reading of a solution of free terminator to 50. With the G-factor set to 1, read 20 L of solution containing 1 L of Terminator Mix and 2 L of 10X Reaction Buffer. Insert the S and P intensities into the following equation and calculate the G-factor for each dye. G = 950S/1050P • The FP value is calculated by following formula: mP = 1000 x [Ivv – Ivh]/[Ivv + Ivh] where Ivv is the emission intensity measured when the excitation and emission polarizers are parallel an

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.

Experts123